DETERMINATION OF INTERNAL RESISTANCE OF A PRIMARY CELL
1. AIM
2. APPARATUS USED
- Potentiometer wire (1-2 meters long)
- Leclanche cell (test cell whose internal resistance is to be determined)
- Standard cell (e.g., Weston cell or another cell with known EMF)
- Galvanometer
- Resistance box (0-100 Ω)
- One-way key
- Two-way key
- Jockey (contact maker)
- Connecting wires
- Ammeter (0-3 A)
- Rheostat
- Battery or power supply (4-6 V)
3. DIAGRAM
4. THEORY
The potentiometer works on the principle that when a constant current flows through a wire of uniform cross-section, the potential difference between any two points on the wire is directly proportional to the length of the wire between those points.
According to Ohm's law, if a current $I$ flows through a resistance $R$, the potential difference $V$ across it is:
In a potentiometer setup, when a cell of EMF $E$ is balanced against a length $l$ of the potentiometer wire, we have:
The internal resistance of a cell is the resistance offered by the cell itself to the flow of current through it. When a current is drawn from a cell, its terminal potential difference $(V)$ is less than its EMF $(E)$ due to the voltage drop across the internal resistance $(r)$:
Where:
- $E$ is the EMF of the cell
- $I$ is the current drawn from the cell
- $r$ is the internal resistance of the cell
- $V$ is the terminal potential difference
In this experiment, we use the potentiometer to measure:
- The EMF $(E)$ of the test cell when no current is drawn (open circuit)
- The terminal potential difference $(V)$ when a current is drawn through a known external resistance
From these measurements, we can calculate the internal resistance of the cell.
5. FORMULA
The internal resistance $(r)$ of the primary cell can be calculated using the formula:
Where:
- $r$ is the internal resistance of the cell
- $R$ is the external resistance connected to the cell
- $E_1$ is the EMF of the cell (measured when no current is drawn)
- $V$ is the terminal potential difference of the cell when current is drawn through resistance $R$
Alternatively, if $l_1$ is the balancing length for EMF and $l_2$ is the balancing length for terminal potential difference, then:
Where $l_1$ and $l_2$ are the balancing lengths on the potentiometer wire.
6. PROCEDURE
-
Setup the circuit:
- Connect the primary circuit as shown in the diagram, consisting of a battery, rheostat, ammeter, and potentiometer wire.
- Connect the secondary circuit with the galvanometer, jockey, and two-way key K₂.
-
Standardize the potentiometer:
- Close the primary circuit using key K₁.
- Adjust the rheostat to ensure sufficient current flows through the potentiometer wire.
- Connect the standard cell E₂ to the circuit using the two-way key K₂.
- Move the jockey along the wire until the galvanometer shows zero deflection. Note this position as the balancing length for the standard cell.
-
Measure the EMF of the test cell:
- Switch the two-way key K₂ to connect the test cell E₁ (without any external resistance).
- Find the balancing point by moving the jockey and note this length as $l_1$.
-
Measure the terminal potential difference:
- Connect a suitable resistance $R$ from the resistance box in series with the test cell.
- Find the new balancing point and note this length as $l_2$.
- Repeat this step with at least 5 different values of $R$.
-
Calculate the internal resistance:
- For each set of readings, calculate the internal resistance using the formula: $r = R(l_1/l_2 - 1)$
- Take the mean value of all calculated internal resistances.
7. OBSERVATION TABLE
Table 1: Standardization of Potentiometer
S.No. | Standard Cell EMF (V) | Balancing Length (cm) | Potentiometer Sensitivity (V/cm) |
---|---|---|---|
1 |
Table 2: Determination of Internal Resistance
S.No. | External Resistance R (Ω) | Balancing Length for EMF $l_1$ (cm) | Balancing Length for Terminal PD $l_2$ (cm) | Internal Resistance $r = R(l_1/l_2 - 1)$ (Ω) |
---|---|---|---|---|
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 |
8. CALCULATIONS
-
Calculate the potentiometer sensitivity:
Sensitivity = \frac{\text{EMF of standard cell}}{\text{Balancing length for standard cell}}
-
For each set of readings, calculate the EMF ($E_1$) and terminal potential difference (V) of the test cell:
E_1 = \text{Sensitivity} \times l_1V = \text{Sensitivity} \times l_2
-
Calculate the internal resistance for each reading using:
r = R\left(\frac{E_1}{V} - 1\right) \text{ or } r = R\left(\frac{l_1}{l_2} - 1\right)
-
Calculate the mean value of the internal resistance:
r_{mean} = \frac{r_1 + r_2 + r_3 + r_4 + r_5}{5}
-
Calculate the percentage error (if an accepted value is available):
\% \text{Error} = \frac{|r_{mean} - r_{accepted}|}{r_{accepted}} \times 100
Sample Calculation:
Given:
- External resistance $R = 5 \, \Omega$
- Balancing length for EMF $l_1 = 80 \, \text{cm}$
- Balancing length for terminal PD $l_2 = 70 \, \text{cm}$
Internal resistance calculation:
9. RESULT
The internal resistance of the given primary cell is _____ Ω ± _____ Ω.
10. PRECAUTIONS
- The potentiometer wire should be of uniform cross-section and free from kinks.
- The connections should be tight and clean to avoid contact resistance.
- The jockey should touch the wire gently to avoid damage to the wire.
- The resistances used should be accurate and in good condition.
- The galvanometer should be sensitive enough to detect small changes in current.
- The battery used in the primary circuit should provide a steady current.
- The rheostat should be adjusted to maintain sufficient current in the potentiometer wire.
- Avoid parallax error while taking readings on the potentiometer scale.
- The room temperature should be noted as EMF varies slightly with temperature.
- The keys should be pressed only when readings are being taken to avoid unnecessary current drain.
11. SOURCES OF ERROR
- Non-uniformity in the cross-section of the potentiometer wire.
- Temperature variations during the experiment affecting the resistance of the wire.
- Contact resistance at junctions and connections.
- Thermo-electric EMF generated at junctions of dissimilar metals.
- Parallax error in reading the position of the jockey.
- Fluctuations in the primary circuit current.
- Zero error in the galvanometer.
- Error in the resistance box values.
- Polarization effects in the cell during the experiment.
- Improper standardization of the potentiometer.